Model | KRY-A15W/3S | KRY-A25W/3S | KRY-A38W/3S | KRY-A60W/3S |
KRY-A15/3S | KRY-A25/3S |
Temp. Range | 0℃~+100℃ |
Temp. control precision | ±0.5℃ | ±0.5℃ | ±0.5℃ | ±0.5℃ |
Temp. feedback | Pt100 | Pt100 | Pt100 | Pt100 |
Temp. display | 0.01k | 0.01k | 0.01k | 0.01k |
Flow control/Group | 2~25 L/min | 2~25 L/min | 2~25 L/min | 1~25 L/min |
Flow control accuracy/Group | ±0.2L | ±0.2L | ±0.2L | ±0.2L |
Number of fluid output groups | 3 Groups | 3 Groups | 3 Groups | 3 Groups |
pressure control/Group | 0.2bar~2.5bar | 0.2bar~2.5bar | 0.2bar~2.5bar | 0.2bar~2.5bar |
flow pressure description | The equipment can control flow and pressure separately. need to control at the same time using different solutions (to be customized) |
Heating power | 10kW | 15kW | 25kW | 38kW |
cooling capacity | 100℃ | 15kW | 25kW | 38kW | 60kW |
20℃ | 15kW | 25kW | 38kW | 60kW |
5℃ | 11kW | 18kW | 27kW | 42kW |
compression | Emerson Valley Wheel Compressor |
expansion | Electronic expansion valve |
oil distributor | Emerson |
Dry filter, Pressure | Danfoss |
Evaporator | plate heat exchanger |
Ciriulating pump | Nangfang centrifugal pump can be selected with imported brand magnetic pump |
Frequency | Delta |
Flow, Pressure sensor | German brand flowmeter, johnson Controls pressure |
input, Display | Siemens 10-inch color touch screen \ Siemens s7-1200 PLC controller |
Communication | Modbus RTU protocol RS485 interface, optional CAN communication bus, Ethernet interface TCP/IP protocol |
Safety protection | With self-diagnosis function; Refrigerator overload protection; High pressure switch, overload relay, thermal protection |
device, low liquid level protection, high temperature protection, sensor fault protection and other safety functions |
refrigerant | R404A/R507C/R125/N40 All four refrigerants are available |
Connection size | There are three groups of inlet and outlet,G3/4 |
Cooling water at 20C° | 3200L/H | 6000L/H | 9000L/H | 15000L/H |
Condenser(W) | Shenshi tube type heat exchanger |
Power 380V50HZ | 22kW | 33kW | 45kW | 68kW |
Dimension cm | 80*120*185 | 100*150*185 | 100*150*185 | 100*150*185 |
Weight | 420kg | 750kg | 1000kg | 1700kg |
Optional | 220V 60HZ three phase 460V 60H three phase |
Optional | The size external circulation pipe line is DN15 or DN20, do not throttle, which will increase the circulation pump load, |
pipe need to be insulated, and adopt rubber and plastic insulation pipe which can afford high temp. |
Optional | When the temperature is lower than the normal temperature, the refrigeration system adopts the refrigeration PID to adjust the cooling capacity, and does not adopt the heating countermeasures, so as to achieve the purpose of energy saving |