KRY and Test Chamber
- Temperature Control Range: -40°C ~ +135°C
- Temperature Control Accuracy: ± 0.5 °C
- Siemens PLC Control Screen And Color Touch Screen
Note: Both Air Cooled And Water Cooled Are Available. Explosion Proof Cabinet Can Be Customized.
Phone: +86-18806176058
Search Search all
Note: Both Air Cooled And Water Cooled Are Available. Explosion Proof Cabinet Can Be Customized.
The liquid refrigerant in the evaporator of the unit is exchanged with the circulating air inside the box, and the temp. rises, from the liquid state to the gaseous state, and is inhaled by the suction port of the compressor.After the compressor compresses the high temperature and high pressure refrigerant gas through the oil separator, the refrigerant oil carried during the compression process is separated clean, then entered into the condenser, and heat is exchanged with the ambient air or cooling water. Change from high temp.and high pressure gas to high pressure normal temp. liquid.The refrigerant liquid at high pressure and normal temp. is changed from high pressure liquid to low pressure liquid through dry filter and then throttled down by expansion valve. Low pressure liquid enters unit evaporator, The circulating air brings the cooling amount to the inside of the box to achieve refrigeration.
In 80L pressure vessel, antifreeze is injected into the vessel and the whole closed circulation system is adopted, thus avoiding the chance of air contact between the antifreeze and the outside world, and ensuring no condensation of water vapor. When the expansion box is protected by nitrogen, the temperature range of the heat conduction liquid will also be widened because its boiling point has been raised. In addition, oxidation from contact with oxygen in the environment is avoided, thus prolonging the service life of the thermal conductive liquid.
The touch interface optimizes the availability of the device. Multi-level users can be implemented through password protection. The specified operator has limited permissions and can only accept settings. This avoids accidental modification and mis-operation of parameters. The operation is continuously monitored. Warnings and failures are clearly displayed and the system automatically shuts down if necessary. Faults can be tracked, so testing does not need to be restarted. At this point, an existing record is invoked.The system has a variety of data and communication channels that can be adjusted according to specific requirements. The temperature control unit can also be remotely controlled via the network or included in a control system Unification.
Model | KRYCH-15W | ||
KRYCH-15 | |||
Temp Range | -40℃~135℃ Anti-freeze fluid | -45℃~100℃ Test chamber | |
Antifreeze flow control | 2L/min~20L/min ±0.2L/min | ||
Antifreeze pressure | Measuring the inlet and outlet pressure and displaying the pressure on the touch screen | ||
Volume of test chamber | 1000*1000*1000 1m³ | ||
Heating power | 7.5kW Anti-freeze fluid | 5kW Test chamber | |
Refrigerating Output Anti-freeze fluid | 135℃ | 7.5kW | |
20℃ | 7.5kW | ||
0℃ | 7.5kW | ||
-20℃ | 4.5kW | ||
-30℃ | 2.5kW | ||
-35℃ | 1.8kW | ||
Refrigerating output Test bin | 100℃ | 6kW | |
-20℃ | 3.5kW | ||
-40℃ | 1.2kW | ||
Temp control precision of antifreeze | ±0.3℃ | ||
High & low temp cavity temp control | ±0.3℃ | ||
Test chamber internal circulating air | 3500m3/h | ||
When the door is opened, the internal circulation fan is automatically closed, and the screen on the touch screen will show that it is in the open state. | |||
Controller | Siemens PLC, Fuzzy PID Control algorithm | ||
Communicating protocol | Ethernet Interface TCP/IP Protocol | ||
Internal temperature feedback | Equipment tank internal temperature, refrigeration system condensation temperature, compressor suction temperature, cooling water temperature (water cooling equipment is configured). | ||
Heat | The maximum heating output power of the system (according to each model); | ||
The heater has triple protection and independent temperature limiter to ensure the safety of heating system | |||
Heating power 15kW adopts regulator, heating power output control adopts 4~20mA linear control. | |||
Refrigerating capacity | Refers to the ability to take away heat at different temperatures (ideal condition), the actual working conditions need to consider environmental heat dissipation, please enlarge properly, and do a good job of insulation measures. | ||
High temp cooling module | The module can be cooled from 150 ℃ at high temperature. | ||
Circulating pump flow | Magnetic driving pump with LNEYA | ||
Source 380V50HZ | 23kW max | ||
Shape dimension | 1250*2350*1900 Same size of air-cooled water without outburst | ||
Compression engine | Emerson Valley Wheel Vortex flexible Compressor | ||
Evaporimeter | Adopting DANFOSS Plate Heat Exchanger / High strength Plate Heat Exchanger | ||
Refrigeration Accessory | Emerson / Danfoss brand dry filter, pressure control, oil separator, etc. | ||
Expansion valve | Danfoss thermal expansion valve / Emerson electronic expansion valve | ||
Operation panel | 7-inch color touch screen, temperature curve display\ EXCEL data export | ||
Safe guarding | With self-diagnosis function; phase sequence phase breaker, freezer overload protection; high-voltage pressure switch, overload relay, thermal protection device and other safety protection functions. | ||
Cryogen | R-404A/R507C | ||
Reserved antifreeze interface inside the box | G3/4 And globe valves | ||
Water-cooled type W | Type W is water cooled | ||
Water-cooled condenser | Double-pipe (heat) exchanger | ||
Cooling water 32℃ | 6m³/H | ||
Shell material | Cold rolled sheet spray (Standard Colour 7035) |
Model | KRYCH-15W | ||
KRYCH-15 | |||
Temp Range | -40℃~135℃ Anti-freeze fluid | -45℃~100℃ Test chamber | |
Antifreeze flow control | 2L/min~20L/min ±0.2L/min | ||
Antifreeze pressure | Measuring the inlet and outlet pressure and displaying the pressure on the touch screen | ||
Volume of test chamber | 1000*1000*1000 1m³ | ||
Heating power | 7.5kW Anti-freeze fluid | 5kW Test chamber | |
Refrigerating Output Anti-freeze fluid | 135℃ | 7.5kW | |
20℃ | 7.5kW | ||
0℃ | 7.5kW | ||
-20℃ | 4.5kW | ||
-30℃ | 2.5kW | ||
-35℃ | 1.8kW | ||
Refrigerating output Test bin | 100℃ | 6kW | |
-20℃ | 3.5kW | ||
-40℃ | 1.2kW | ||
Temp control precision of antifreeze | ±0.3℃ | ||
High & low temp cavity temp control | ±0.3℃ | ||
Test chamber internal circulating air | 3500m3/h | ||
When the door is opened, the internal circulation fan is automatically closed, and the screen on the touch screen will show that it is in the open state. | |||
Controller | Siemens PLC, Fuzzy PID Control algorithm | ||
Communicating protocol | Ethernet Interface TCP/IP Protocol | ||
Internal temperature feedback | Equipment tank internal temperature, refrigeration system condensation temperature, compressor suction temperature, cooling water temperature (water cooling equipment is configured). | ||
Heat | The maximum heating output power of the system (according to each model); | ||
The heater has triple protection and independent temperature limiter to ensure the safety of heating system | |||
Heating power 15kW adopts regulator, heating power output control adopts 4~20mA linear control. | |||
Refrigerating capacity | Refers to the ability to take away heat at different temperatures (ideal condition), the actual working conditions need to consider environmental heat dissipation, please enlarge properly, and do a good job of insulation measures. | ||
High temp cooling module | The module can be cooled from 150 ℃ at high temperature. | ||
Circulating pump flow | Magnetic driving pump with LNEYA | ||
Source 380V50HZ | 23kW max | ||
Shape dimension | 1250*2350*1900 Same size of air-cooled water without outburst | ||
Compression engine | Emerson Valley Wheel Vortex flexible Compressor | ||
Evaporimeter | Adopting DANFOSS Plate Heat Exchanger / High strength Plate Heat Exchanger | ||
Refrigeration Accessory | Emerson / Danfoss brand dry filter, pressure control, oil separator, etc. | ||
Expansion valve | Danfoss thermal expansion valve / Emerson electronic expansion valve | ||
Operation panel | 7-inch color touch screen, temperature curve display\ EXCEL data export | ||
Safe guarding | With self-diagnosis function; phase sequence phase breaker, freezer overload protection; high-voltage pressure switch, overload relay, thermal protection device and other safety protection functions. | ||
Cryogen | R-404A/R507C | ||
Reserved antifreeze interface inside the box | G3/4 And globe valves | ||
Water-cooled type W | Type W is water cooled | ||
Water-cooled condenser | Double-pipe (heat) exchanger | ||
Cooling water 32℃ | 6m³/H | ||
Shell material | Cold rolled sheet spray (Standard Colour 7035) |